Identificação de fatores que alteram a confiabilidade da temperatura axilar em relação à temperatura de cateter de artéria pulmonar
Identification of factors that alter axillary temperature reliability in relation to pulmonary artery catheter temperature
Identificación de factores que alteran la confiabilidad de la temperatura axilar en relación a la temperatura del catéter de la arteria pulmonar

REME rev. min. enferm; 23 (), 2019
Publication year: 2019

OBJETIVOS:

identificar fatores que alteram a confiabilidade da temperatura axilar em relação à temperatura de cateter artéria pulmonar.

MÉTODOS:

estudo de medidas repetidas utilizando 67 aferições de temperatura axilar coletadas de 24 pacientes internados em centros de terapia intensiva de dois hospitais da região metropolitana de Belo Horizonte entre 2017 e 2018. Realizada análise descritiva e regressão linear dos dados.

RESULTADOS:

foram encontradas acurácia de 0,48 e precisão de 0,47 referentes à temperatura axilar aferida com termômetro digital comparada com a temperatura de cateter de artéria pulmonar. Os fatores que alteram a confiabilidade da temperatura axilar foram o índice de massa corporal e a dose de dobutamina.

CONCLUSÃO:

os fatores encontrados foram inéditos em relação à alteração da confiabilidade da temperatura axilar e poderão auxiliar enfermeiros na tomada de decisão ao escolherem um método mais preciso para estimativa da temperatura real do corpo. Ainda é necessária a realização de estudos com amostragem maior para avaliação dos fatores intervenientes da confiabilidade de técnicas de temperatura não invasivas, como a temperatura axilar.(AU)

Objectives:

to identify factors that alter the reliability of axillary temperature in relation to pulmonary artery catheter temperature.

Methods:

studying repeated measures using 67 axillary temperature measurements collected from 24 patients admitted to intensive care centers of two hospitals in the metropolitan region of Belo Horizonte between 2017 and 2018. Descriptive analysis and linear regression of the data were performed.

Results:

accuracy of 0.48 and precision of 0.47 were found for axillary temperature measured with a digital thermometer compared with pulmonary artery catheter temperature. Factors that change the reliability of axillary temperature were body mass index and dobutamine dose.

Conclusion:

the found factors were unprecedented in relation to the change in axillary temperature reliability and may help nurses in their decision making by choosing a more accurate method to estimate the actual body temperature. Studies with larger sampling are still necessary to evaluate the intervening factors of the reliability of noninvasive temperature techniques, such as axillary temperature.(AU)

Objetivos:

identificar los factores que alteran la fiabilidad de la temperatura axilar en relación con la temperatura del catéter de la arteria pulmonar.

Métodos:

estudio de mediciones repetidas utilizando 67 mediciones de temperatura axilar de 24 pacientes ingresados en centros de cuidados intensivos de dos hospitales de la región metropolitana de Belo Horizonte entre 2017 y 2018. Análisis descriptivo y regresión lineal de los datos.

Resultados:

se encontró precisión de 0,48 y precisión de 0,47 para la temperatura axilar medida con termómetro digital en comparación a la temperatura del catéter de la arteria pulmonar. Los factores que alteran la fiabilidad de la temperatura axilar fueron el índice de masa corporal y la dosis de dobutamina.

Conclusión:

los factores encontrados no tienen precedentes en relación con la alteración en la confiabilidad de la temperatura axilar y pueden ayudar a los enfermeros en la toma de decisiones al elegir un método más preciso para estimar la temperatura corporal real. Son necesarios estudios con muestras más amplias para evaluar los factores que intervienen en la fiabilidad de las técnicas de temperatura no invasivas, como la temperatura axilar.(AU)