Objetivo:
 identificar as causas da fadiga de alarmes em estudos de usabilidade de bomba de infusão em terapia intensiva pediátrica. Métodos:
 Realizou-se uma revisão integrativa nas bases de dados LILACS, SciELO, IBECS, SCOPUS e MEDLINE. Resultados:
 Foram identificadas 1.164 publicações e selecionados seis estudos primários que emergiram em duas temáticas:
 compreender as causas dos alarmes da bomba de infusão que constou: biblioteca de fármacos incompleta, limites absolutos e relativos rígidos, falta de protocolo de preparo e administração de medicação, período de férias dos funcionários; e as medidas que podem evitar a fadiga de seus alarmes.Conclusão:
falhas na usabilidade de bombas de infusão aumentam os alarmes desnecessários que podem promovem sua fadiga. A utilização segura de bombas de infusão demanda uma equipe que monitore suas práticas e atue promovendo mudanças no contexto de trabalho                                        
                                                                                                                    
                                            
Objective:
 to identify the causes of alarm fatigue in studies of infusion pump usability in pediatric intensive care. Methods:
 an integrative review was carried out in the LILACS, SciELO, IBECS, SCOPUS and MEDLINE databases. Results:
 1,164 publications were identified and six primary studies were selected that emerged in two themes: understanding the causes of the infusion pump alarms that consisted of: incomplete drug library, absolute and strict relative limits, lack of preparation protocol and medication administration, employee vacation period;  and  measures  that  can  prevent  fatigue  from  your  alarms.  Conclusion:
 the  causes  of  alarm  fatigue  involve  low  user  interaction with  the  equipment,  inadequate  work  processes  and  low  investment  in  preventive  measures  for  its  occurrence  The  safe  use  of  infusion  pumps  requires a team to monitor their practices and act by promoting changes in the work context                                        
                                                                                                                    
                                            
Objetivo:
 identificar  las  causas  de  la  fatiga  de  alarma  en  estudios  de  usabilidad  de  bombas  de  infusión  en  cuidados  intensivos  pediátricos.  Métodos:
 se realizó una revisión integradora en las bases de datos LILACS, SciELO, IBECS, SCOPUS y MEDLINE. Resultados:
 se identificaron 1.164 publicaciones y se seleccionaron seis estudios primarios que surgieron en dos temas:
 comprender las causas de las alarmas de la bomba de infusión que consistían en: biblioteca de medicamentos incompleta, límites relativos absolutos y estrictos, falta de protocolo de preparación y administración de  medicamentos,  período  de  vacaciones  de  los  empleados;  y  medidas  que pueden prevenir la fatiga de sus alarmas. Conclusión:
 las causas de la fatiga de las alarmas involucran baja interacción del usuario con el equipo, procesos de trabajo inadecuados y baja inversión en medidas preventivas para su ocurrencia. El uso seguro de las bombas de infusión requiere que un  equipo  monitoree  sus  prácticas  y  actúe  promoviendo  cambios  en  el  contexto de trabajo