Noise level analysis in adult intensive care unit
Análise do nível de ruído em unidade de terapia intensiva adulto

Rev Rene (Online); 17 (4), 2016
Publication year: 2016

Objective:

to analyze the noise level in adult intensive care unit.

Methods:

a quantitative study, in which the sound levels of the intensive care unit have been assessed by means of a decibel meter.

Results:

comparing the groups, there was a reduction in noise levels in both periods studied, but only in the afternoon there was a statistically significant difference (p<0.05). The health professionals pointed out that the unit had moderate noise, coming mainly from equipment and professionals.

Conclusion:

adjusting the ventilator alarms contributed to the reduction of noise levels in the unit, and there was the perception that it is a moderate noise environment, although the noise levels in decibels observed were above the recommended values.
Analisar o nível de ruído em unidade de terapia intensiva adulto.

Métodos:

estudo quantitativo, em que se aferiram os níveis sonoros da unidade de terapia intensiva por meio de decibelímetro.

Resultados:

comparando-se os grupos, verificou-se a redução dos níveis de ruído em ambos os períodos estudados, porém somente no período da tarde houve diferença estatisticamente significativa (p<0,05). Os profissionais apontaram que a unidade apresentava ruído moderado, oriundo principalmente dos equipamentos e profissionais.

Conclusão:

o ajuste dos alarmes de ventilação mecânica contribuiu para a redução dos níveis de ruído na unidade, e houve percepção de que se trata de um ambiente de ruído moderado, apesar de os níveis de ruído em decibéis observados estarem acima dos valores recomendados.